AXIS-NV: High Throughput Optical Streak Camera

AXIS-NV is used for time-resolved spectroscopy of subnanosecond events
in the visible and IR range.

Its wide aperture allows to streak a 1000 point image with a time resolution
of 10-50 picosecond with an extremely high dynamic range.

Applications:

  • High Energy Physics
  • Z-Pinch Science
  • Plasma Diagnostics

Specifications Overview


Streak tube

Photonis P500 family:
These sealed tubes provide an excellent spatial resolution and an extremely high dynamic range.

Spectral Response

200-850 nm (U curve) or 300-1400 nm (I curve)

Click on image to enlarge

Spatial Resolution

Photocathode length (X) 35 mm
Photocathode effective width (Y) adjustable from 50µm to 4mm
Spatial Resolution (dx) 60µm to 30µm with 50% contrast
Number of spatial resolution units in screen (X/dx) 500 to 1000 points

Single-shot Sweep

Available Ranges (T) 2 ns to 1 ms
Number of time-resolution units in screen (Nt) 600 in the narrow slit mode
Single-shot Time Resolution 5 ps to 150ps

Averaged Sweep

Trigger type Electrical (10-35V in 50Ω)
Trigger rate single shot to 10 kHz
Trigger jitter <15 ps RMS

Readout

Readout type Digital CCD camera
CCD chip size 4008 x 2672 pixels
Digitizer 14 bits
Coupling to streak tube 1.38:1 Fiber optic taper
Effctive detection area 50 mm (time axis) x 33 mm (space axis)
Cooling air

Operation modes

Normal sweep The sweep starts out of the screen, crosses the whole screen and ends outside the screen.
Timing mode The sweep starts in the screen and ends also the screen. It is used to synchronize the streak camera to the event.
Focus mode A static voltage is applied to position the trace close to the center of the screen. The sweep trigger is disabled. This mode is used to adjust the incident light into the streak camera.

General

Electrical input Universal AC, 110-240V, 50-60 Hz
Certification CE
Tests and Calibration Before shipping, each system is tested on a femtosecond laser at the Advanced Laser Light Source (Canada)
Installation and Training A qualified engineer is sent to your laboratory to install the system and train users

Options

Blanking circuits With the blanking option, streak tube is blocked during the time the signals sweeps back to its start position. This feature is required when the optical signal lasts longer than the sweep range.
Liquid chiller Liquid chiller and plumbing to cool the CCD camera.
Remote control over Optical Ethernet Web-based user interface to operate system via optical Ethernet 1 Gb/s (1000 Base SX) on 2 SMA 905 connectors (or other common fiber optics standards).
Custom input optics Custom designed optical setup that accommodate special needs.
Slow scan mode Operation mode in which the sweep crosses the whole screen in 0.5 second. It is used to characterize spatial non-uniformity in the instrumental response by illuminating with a constant brightness light source.
24V supply To power the whole system with a DC voltage between 23V and 28V.
Spectrometer Custom spectrometer that fits special needs.